Answers are grounded in this paper and cite their sources.
Measurement of plasma osmolality (Posm) remains popular for assessing hydration status in exercise science. However, a controlled reliability assessment of micro-osmometry using small sample volumes to measure Posm remains to be performed. This study aimed to examine the reliability of a cryoscopic micro-osmometer requiring 15-μL samples across varied environmental conditions. Thirteen young adults (26.4 ± 5.7 years) completed a 20-min steady-state cycling protocol on separate occasions in cool/dry conditions and hot/humid conditions. Posm was assessed using fingertip capillary samples taken in duplication prior to and following 20-min acclimatization and recovery periods around the cycling bout. Absolute (typical error of measurement = .62–.83 mOsmol·kg–1; coefficient of variation = .9% to 1.3%) and relative (intraclass correlation coefficient = .60–.85) reliability of this approach were supported. These findings suggest micro-osmometry requiring small plasma samples (15 μL) can be reliably used by practitioners to assess longitudinal changes and cross-sectional hydration status in varied environmental conditions.
No takes yet. Share an insight, caveat, or question.
Scanlan et al. (2016) studied this question.