Key result
A bipolar pacing electrode was successfully implanted without interference from a concomitant vertebral nerve stimulator, whereas a unipolar system registered 0.25 mV potentials.
Case Report (n=1)
Bipolar pacing electrodes may prevent electromagnetic interference from concomitant nerve stimulators that is otherwise observed with unipolar systems.
Answers are grounded in this paper and cite their sources.
Bipolar pacing may prevent nerve stimulator interference; hypothesis-generating from a single case and should not yet change practice.
Electromagnetic inference is a recognized difficulty in concomitant cardiac pacing and nerve stimulation. A case is reported in which a vertebral nerve stimulator was implanted, with a transmitter antenna distal to the right nipple. Six months later, intermittent S-A block with Adams-Stokes attacks urgently necessitated cardiac pacing. The ECG lead II showed nerve stimulator potentials of 0.75 mV. Intracardiac electrogram with unipolar electrode registered potential of 0.25 mV, but with a bipolar system no such potentials were observed. A bipolar pacing electrode was therefore implanted, with an interference from the nerve stiumluator. Various aspects of interference are discussed.
No takes yet. Share an insight, caveat, or question.
Sylvén et al. (1979) conducted a case report in Intermittent S-A block with Adams-Stokes attacks concomitant with vertebral nerve stimulator (n=1). Bipolar pacing electrode vs. Unipolar electrode was evaluated on Electromagnetic interference from the nerve stimulator. A bipolar pacing electrode was successfully implanted without interference from a concomitant vertebral nerve stimulator, whereas a unipolar system registered 0.25 mV potentials.
Synapse has enriched 10 related papers. Consider them for comparative context: