Key result
Heart rate detection using an electronic weighing scale based on ballistocardiographic force signals estimated RR time intervals with an error of +/-21 ms compared to ECG.
Why the study?
Does heart rate detection using ballistocardiographic force signals from an electronic weighing scale accurately measure RR intervals compared to ECG in volunteers?
Does heart rate detection using ballistocardiographic force signals from an electronic weighing scale accurately measure RR intervals compared to ECG in volunteers?
Answers are grounded in this paper and cite their sources.
A novel technique using an electronic weighing scale to detect ballistocardiographic force signals can accurately estimate heart rate for short-term monitoring.
We propose a novel technique for beat-to-beat heart rate detection based on the ballistocardiographic (BCG) force signal from a subject standing on a common electronic weighing scale. The detection relies on sensing force variations related to the blood acceleration in the aorta, works even if wearing footwear and does not require any sensors attached to the body because it uses the load cells in the scale. We have devised an approach to estimate the sensitivity and frequency response of three commercial weighing scales to assess their capability to detect the BCG force signal. Static sensitivities ranged from 490 nV V(-1) N(-1) to 1670 nV V(-1) N(-1). The frequency response depended on the subject's mass but it was broad enough for heart rate estimation. We have designed an electronic pulse detection system based on off-the-shelf integrated circuits to sense heart-beat-related force variations of about 0.24 N. The signal-to-noise ratio of the main peaks of the force signal detected was higher than 30 dB. A Bland-Altman plot was used to compare the RR time intervals estimated from the ECG and BCG force signals for 17 volunteers. The error was +/-21 ms, which makes the proposed technique suitable for short-term monitoring of the heart rate.
No takes yet. Share an insight, caveat, or question.
González-Landaeta et al. (2008) studied Heart rate detection (n=17). Electronic weighing scale (BCG force signal) vs. Electrocardiogram (ECG) was evaluated on RR time intervals (+/-21 ms error). Heart rate detection using an electronic weighing scale based on ballistocardiographic force signals estimated RR time intervals with an error of +/-21 ms compared to ECG.
Synapse has enriched 10 related papers. Consider them for comparative context: