Non-constrained monitoring of systolic blood pressure on a weighing scale

View Full Paper
JSJae Hyuk ShinKLKang Moo LeeKPKwang Suk Park

Key Result

A novel cuffless blood pressure estimation technique using ECG and BCG measured on a weighing scale showed reliable correlation with reference systolic blood pressure measurements.

Structured PICO

Does a novel technique using ECG and BCG on a weighing scale accurately estimate systolic blood pressure compared to a reference system?

P
Population
Human subjects (sample size and demographics not specified in abstract)
I
Intervention
Non-constrained and cuffless blood pressure estimation based on electrocardiogram (ECG) and ballistocardiogram (BCG) measured using an electronic weighing scale
C
Comparator
Finapres (non-invasive blood pressure measurement system) as a reference
O
Outcome
Accuracy and correlation of estimated systolic blood pressure (SBP) compared to measured SBPsurrogate

A novel technique using ECG and BCG measured on a weighing scale provides reliable estimates of systolic blood pressure compared to a reference system, enabling potential home healthcare monitoring.

Abstract

In this study, we developed a novel technique for estimating non-constrained and cuffless blood pressure (BP) that was based on electrocardiogram (ECG) and ballistocardiogram (BCG). The BCG was non-invasively measured using a common electronic weighing scale when a subject was standing on it. The ECG was measured using three different methods: on the chest using Ag/AgCl electrodes, on the hands using dry electrodes and on the feet also using dry electrodes. For a BP correlated parameter, a time interval parameter, which was defined as the time difference between the ECG R-peak and BCG J-peak, was employed for evaluating and estimating beat-to-beat BP. Under a BP varying experiment with a Valsalva manoeuvre, the R-J intervals were extracted at every beat cycle and a systolic blood pressure (SBP) estimation equation was established using linear regression analysis for each subject. In the case of feet delivered ECG (F-ECG), an ensemble average technique synchronized at the BCG J-peak point was applied to extract the ECG signal from the feet. The performance of the proposed method was evaluated using Finapres, a non-invasive blood pressure measurement system, as a reference BP signal, and a scatter plot was used to find the regression line between the reference values and estimated BPs. A moving-window averaging technique was applied to remove the high-frequency noise in the R-J intervals and was applied to enhance the accuracy of the SBP estimation. For all individuals, the estimated SBP was similar to the measured SBP with a reliable correlation, which makes the proposed method suitable for use in a home healthcare system to monitor blood pressure on a weighing scale at the same time as measuring weight.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Shin et al. (2009) studied Blood pressure monitoring. Cuffless blood pressure estimation using ECG and BCG on a weighing scale vs. Finapres non-invasive blood pressure measurement system was evaluated on Systolic blood pressure estimation accuracy. A novel cuffless blood pressure estimation technique using ECG and BCG measured on a weighing scale showed reliable correlation with reference systolic blood pressure measurements.