Key result
Novel BCG and IPG weighing scale estimates systolic BP within ~7 mmHg error, achieving grade C.
Why the study?
The goal was to develop a minimal contact blood pressure measurement method based on a commercial body weight-fat scale to support daily eHealth monitoring.
Does a minimal contact blood pressure measurement system using ballistocardiogram and impedance plethysmogram signals from a commercial body weight-fat scale accurately estimate systolic and diastolic blood pressure compared to standard cuff-based and ECG-PPG cuffless methods in young adults?
Does a minimal contact blood pressure measurement system using ballistocardiogram and impedance plethysmogram signals from a commercial body weight-fat scale accurately estimate systolic and diastolic blood pressure compared to standard cuff-based and ECG-PPG cuffless methods in young adults?
A modified commercial body weight-fat scale extracting BCG and IPG signals can estimate blood pressure without body-worn sensors to an IEEE Grade C accuracy level, offering a convenient approach for daily eHealth trend monitoring.
Answers are grounded in this paper and cite their sources.
May enable convenient daily BP trend monitoring via scales; leaves open validation needs before clinical or research adoption.
Electronic health (eHealth) is a strategy to improve the physical and mental condition of a human, collecting daily physiological data and information from digital apparatuses. Body weight and blood pressure (BP) are the most popular and important physiological data. The goal of this study is to develop a minimal contact BP measurement method based on a commercial body weight-fat scale, capturing biometrics when users stand on it. The pulse transit time (PTT) is extracted from the ballistocardiogram (BCG) and impedance plethysmogram (IPG), measured by four strain gauges and four footpads of a commercial body weight-fat scale. Cuffless BP measurement using the electrocardiogram (ECG) and photoplethysmogram (PPG) serves as the reference method. The BP measured by a commercial BP monitor is considered the ground truth. Twenty subjects participated in this study. By the proposed model, the root-mean-square errors and correlation coefficients (r2s) of estimated systolic blood pressure and diastolic blood pressure are 7.3 ± 2.1 mmHg and 4.5 ± 1.8 mmHg, and 0.570 ± 0.205 and 0.284 ± 0.166, respectively. This accuracy level achieves the C grade of the corresponding IEEE standard. Thus, the proposed method has the potential benefit for eHealth monitoring in daily application.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2023) studied Healthy young adults undergoing blood pressure measurement (n=20). Minimal contact cuffless blood pressure measurement via commercial body weight-fat scale (BCG and IPG) vs. Reference cuffless BP measurement (ECG and PPG) and ground-truth commercial oscillometric cuff BP monitor (Omron HM-7320) was evaluated on Root-mean-square error (E_RMS) and correlation coefficient (r^2) of estimated systolic blood pressure (SBP) using Model 4 (Equation 6). Ballistocardiogram and impedance plethysmogram scale measurements estimated systolic and diastolic BP with root-mean-square errors of 7.2 ± 2.2 mmHg and 4.4 ± 1.9 mmHg, achieving an IEEE grade C.
Synapse has enriched 10 related papers. Consider them for comparative context: