Key result
Scale-based BCG and toe PPG estimate systolic blood pressure within ~7 mmHg root-mean-square error.
Why the study?
Developing touchless blood pressure measurement offers potential benefits for reducing the risk of COVID-19 infection.
Can a cuffless method combining weight-scale ballistocardiography and toe photoplethysmography accurately estimate blood pressure compared to standard oscillometric cuff and ECG-finger PPG reference measurements in young adults?
Observational (n=20)
No
Can a cuffless method combining weight-scale ballistocardiography and toe photoplethysmography accurately estimate blood pressure compared to standard oscillometric cuff and ECG-finger PPG reference measurements in young adults?
This proof-of-concept study demonstrates that blood pressure can be estimated without a cuff by combining ballistocardiography from a commercial weight scale with toe photoplethysmography, achieving root-mean-square errors of 6.7 mmHg for systolic and 4.8 mmHg for diastolic blood pressure.
Answers are grounded in this paper and cite their sources.
Supports scale-based cuffless BP estimation development; leaves open clinical validation in larger, diverse populations.
Currently, in terms of reducing the infection risk of the COVID-19 virus spreading all over the world, the development of touchless blood pressure (BP) measurement has potential benefits. The pulse transit time (PTT) has a high relation with BP, which can be measured by electrocardiogram (ECG) and photoplethysmogram (PPG). The ballistocardiogram (BCG) reflects the mechanical vibration (or displacement) caused by the heart contraction/relaxation (or heart beating), which can be measured from multiple degrees of the body. The goal of this study is to develop a cuffless and touchless BP-measurement method based on a commercial weight scale combined with a PPG sensor when measuring body weight. The proposed method was that the PTTBCG-PPGT was extracted from the BCG signal measured by a weight scale, and the PPG signal was measured from the PPG probe placed at the toe. Four PTT models were used to estimate BP. The reference method was the PTTECG-PPGF extracted from the ECG signal and PPG signal measured from the PPG probe placed at the finger. The standard BP was measured by an electronic blood pressure monitor. Twenty subjects were recruited in this study. By the proposed method, the root-mean-square error (ERMS) of estimated systolic blood pressure (SBP) and diastolic blood pressure (DBP) are 6.7 ± 1.60 mmHg and 4.8 ± 1.47 mmHg, respectively. The correlation coefficients, r2, of the proposed model for the SBP and DBP are 0.606 ± 0.142 and 0.284 ± 0.166, respectively. The results show that the proposed method can serve for cuffless and touchless BP measurement.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2022) conducted an observational in Healthy young adults undergoing exercise-induced blood pressure elevation (n=20). Cuffless and touchless blood pressure estimation using weight-scale ballistocardiogram (BCG) and toe photoplethysmogram (PPG) pulse transit time (PTT) vs. Reference PTT method using Lead I ECG and finger PPG, compared against a standard oscillometric electronic blood pressure monitor was evaluated on Root-mean-square error (ERMS) and correlation coefficient (r2) of estimated systolic blood pressure (SBP) using the inverse PTT model (Equation 4) (ERMS 6.7 ± 1.60 mmHg (r2 = 0.606 ± 0.142) vs 5.4 ± 1.71 mmHg (r2 = 0.742 ± 0.131)). Weight-scale BCG and toe PPG pulse transit time estimated systolic and diastolic blood pressure with root-mean-square errors of 6.7 ± 1.60 mmHg and 4.8 ± 1.47 mmHg.