Key result
Attended and unattended home BP measurements show comparable diagnostic accuracy but low sensitivity versus 24-hour ABPM.
Why the study?
Community health workers increasingly screen for hypertension in rural sub-Saharan Africa, but the diagnostic accuracy of attended versus unattended automated blood pressure measurement at home had not been compared.
Does attended versus unattended automated blood pressure measurement by community health workers have comparable diagnostic accuracy to 24-hour ambulatory blood pressure monitoring in rural sub-Saharan African adults?
Population
1220 individuals aged >= 30 years in Kenya and The Gambia
Comparison
Attended vs unattended automated blood pressure vs 24-hour ABPM
Design
Cross-sectional, population-based diagnostic accuracy study
Authors
Loading...
Attended and unattended measurements show comparable accuracy with high specificity but low sensitivity; supports equivalence but leaves open screening utility pending prospective validation.
Cross-Sectional (n=1,220)
Index and reference tests blinded to each other
Randomly assigned order of BP measurements
Yes
Does attended versus unattended automated blood pressure measurement by community health workers have comparable diagnostic accuracy to 24-hour ambulatory blood pressure monitoring in rural sub-Saharan African adults?
Attended and unattended automated blood pressure measurements by community health workers have comparable diagnostic accuracy with high specificity but low sensitivity compared to 24-hour ambulatory monitoring, suggesting they are useful for initial screening but require confirmatory testing.
Lucinde et al. (2026) conducted a cross-sectional in Hypertension (n=1,220). Attended and unattended automated home blood pressure measurements vs. 24-hour ambulatory blood pressure monitoring (ABPM) was evaluated on Diagnostic accuracy (Area under the receiver operating characteristic curve). Attended and unattended automated home blood pressure measurements conducted by community health workers had comparable diagnostic accuracy (AUC 0.76 vs 0.74 in Kenya; 0.68 vs 0.68 in The Gambia) but low sensitivity compared to 24-hour ABPM.