Key result
Two-stage percutaneous closure of severe paravalvular leak using Amplatzer vascular plug 2 was successful in a patient with a stentless aortic bioprosthesis, remaining stable at 12 months.
Case Report (n=1)
A two-stage percutaneous closure using AVP 2 is a feasible and effective strategy for managing severe paravalvular leak in patients with stentless aortic bioprostheses without causing prosthesis compression.
Answers are grounded in this paper and cite their sources.
Supports staged AVP 2 closure for paravalvular leak in stentless bioprostheses; hypothesis-generating and requires prospective validation.
We report a case of successful 2-stage percutaneous closure of severe paravalvular leak (PVL) in a patient with stentless aortic bioprosthesis. Threat of pliable prosthesis compression and long course of PVL posed main considerations. Accordingly, Amplatzer vascular plug (AVP) 2 was chosen as occluder. It was delivered transfemorally under fluoroscopy and transesophageal echocardiography guidance. No prosthesis distortion was provoked at the expense of moderate residual leak. Its symptoms after 3 months spurred another AVP 2 implantation, which lead to PVL closure with no effect on valve area. Marked reduction of left ventricle cavity size and NT-proBNP plasma concentration was promptly noted and remains stable in 12-month follow-up.
No takes yet. Share an insight, caveat, or question.
Smołka et al. (2012) conducted a case report in severe paravalvular leak (PVL) (n=1). 2-stage percutaneous closure using Amplatzer vascular plug (AVP) 2 was evaluated on PVL closure. Two-stage percutaneous closure of severe paravalvular leak using Amplatzer vascular plug 2 was successful in a patient with a stentless aortic bioprosthesis, remaining stable at 12 months.
Synapse has enriched 10 related papers. Consider them for comparative context: