Answers are grounded in this paper and cite their sources.
Case series demonstrates real-time echocardiographic guidance improves procedural safety during balloon atrial septostomy in transposition of the great arteries, indicating reduced trauma risk.
Two dimensional echocardiography has been used to guide balloon atrial septostomy in five cases of transposition of the great arteries. A single lumen balloon catheter was used. The catheter can be accurately located within the left atrium before inflation of the balloon, thus avoiding some of the traumatic accidents possible during septostomy. The balloon can be observed creating the atrial defect and the approximate size of the resulting defect noted. It can also be seen if the foramen ovale is only being stretched and not torn. The use of the two dimensional echocardiogram during septostomy should significantly contribute to the speed, efficiency, and safety of the procedure.
No takes yet. Share an insight, caveat, or question.
Allan et al. (1982) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: