Key result
B-mode ultrasound imaging yielded significantly different cIMT measurements compared to M-mode (mean difference 0.03 mm; 95% CI 0.02-0.05; p<0.001), and showed higher inter-observer reproducibility.
Why the study?
Common carotid artery intima-media thickness is associated with cardiovascular disease, but few studies have evaluated the variability of different ultrasound imaging modes for its measurement.
Does B-mode ultrasound imaging compared to M-mode ultrasound imaging improve inter-observer reproducibility in measuring cIMT in healthy subjects?
Observational
Does B-mode ultrasound imaging compared to M-mode ultrasound imaging improve inter-observer reproducibility in measuring cIMT in healthy subjects?
Mean Difference: 0.03 (95% CI 0.02–0.05)
p-value: p=<0.001
B-mode ultrasound demonstrates higher inter-observer reproducibility than M-mode for measuring common carotid intima-media thickness, though there is poor agreement between the two modes.
B-mode may improve cIMT reproducibility in research; leaves open standardization versus M-mode for clinical use.
Background: Common carotid artery intima-media thickness (cIMT) has been associated with cardiovascular diseases, including stroke and coronary heart disease. There is a limited number of studies that specifically address the variability of ultrasound imaging modes in measuring cIMT. Therefore, this study aimed to assess the agreement-level and inter-observer reproducibility of cIMT measurements using B-mode and M-mode. Materials and Methods: B-mode and M-mode ultrasound imaging were used in measuring cIMT of healthy subjects using linear-array transducer. Inter-imaging mode agreement and inter-observer reproducibility of B-mode and M-mode were assessed using Intraclass correlation coefficient (ICC). Differences in cIMT measurements between imaging modes and observers were determined using paired t-test Results: There was significant difference between B-mode and M-mode in measuring cIMT (mean difference (MD), 0.03 mm, 95% CI 0.02 - 0.05, p<0.001). Intra-imaging mode agreement was poor with ICC of 0.39. Inter-observer reproducibility of cIMT measured using B-mode and M-mode were good and moderate, respectively, (B-mode ICC of 0.76, 95% CI 0.64 - 0.84, p<0.01; M-mode ICC of 0.68, 95% CI 0.52 - 0.78, p<0.01). No significant difference between the two observers in measuring cIMT measured using B-mode or M-mode (B-mode p=0.3; M-mode p=0.4). Conclusion: There was a significant difference and poor intra-imaging mode agreement between B-mode and M-mode in measuring cIMT. B-mode showed higher inter-observer reproducibility of cIMT measurement than M-mode. Further studies investigating the variability and agreement between ultrasound imaging modes of patients with thickened cIMT are required.
No takes yet. Share an insight, caveat, or question.
Alghamdi et al. (2023) conducted an observational in Healthy subjects. B-mode ultrasound imaging vs. M-mode ultrasound imaging was evaluated on Difference in cIMT measurements between imaging modes (MD 0.03 mm, 95% CI 0.02 - 0.05, p=<0.001). B-mode ultrasound imaging yielded significantly different cIMT measurements compared to M-mode (mean difference 0.03 mm; 95% CI 0.02-0.05; p<0.001), and showed higher inter-observer reproducibility.