Key result
B-mode ultrasound yields higher reproducibility than M-mode for cIMT despite poor inter-mode agreement.
Why the study?
Common carotid artery intima-media thickness is associated with cardiovascular diseases, but a limited number of studies specifically address the variability and agreement of B-mode and M-mode ultrasound imaging in measuring it.
Does B-mode ultrasound provide better inter-observer reproducibility and agreement compared to M-mode ultrasound for manual measurement of common carotid artery intima-media thickness in healthy subjects?
Cross-Sectional (n=50)
No
Does B-mode ultrasound provide better inter-observer reproducibility and agreement compared to M-mode ultrasound for manual measurement of common carotid artery intima-media thickness in healthy subjects?
Manual cIMT measurements differ significantly and show poor agreement between B-mode and M-mode ultrasound, with B-mode demonstrating superior inter-observer reproducibility, indicating that clinicians should consistently use B-mode for reliable cIMT assessment.
Answers are grounded in this paper and cite their sources.
B-mode and M-mode yield differing cIMT values with poor agreement; leaves open mode standardization for reproducible measurements in studies.
Background: Common carotid artery intima-media thickness (cIMT) has been associated with cardiovascular diseases, including stroke and coronary heart disease. There is a limited number of studies that specifically address the variability of ultrasound imaging modes in measuring cIMT. Therefore, this study aimed to assess the agreement-level and inter-observer reproducibility of cIMT measurements using B-mode and M-mode. Materials and Methods: B-mode and M-mode ultrasound imaging were used in measuring cIMT of healthy subjects using linear-array transducer. Inter-imaging mode agreement and inter-observer reproducibility of B-mode and M-mode were assessed using Intraclass correlation coefficient (ICC). Differences in cIMT measurements between imaging modes and observers were determined using paired t-test Results: There was significant difference between B-mode and M-mode in measuring cIMT (mean difference (MD), 0.03 mm, 95% CI 0.02 - 0.05, p<0.001). Intra-imaging mode agreement was poor with ICC of 0.39. Inter-observer reproducibility of cIMT measured using B-mode and M-mode were good and moderate, respectively, (B-mode ICC of 0.76, 95% CI 0.64 - 0.84, p<0.01; M-mode ICC of 0.68, 95% CI 0.52 - 0.78, p<0.01). No significant difference between the two observers in measuring cIMT measured using B-mode or M-mode (B-mode p=0.3; M-mode p=0.4). Conclusion: There was a significant difference and poor intra-imaging mode agreement between B-mode and M-mode in measuring cIMT. B-mode showed higher inter-observer reproducibility of cIMT measurement than M-mode. Further studies investigating the variability and agreement between ultrasound imaging modes of patients with thickened cIMT are required.
No takes yet. Share an insight, caveat, or question.
Alghamdi et al. (2023) conducted a cross-sectional in Healthy subjects undergoing common carotid artery intima-media thickness (cIMT) assessment (n=50). B-mode ultrasound imaging vs. M-mode ultrasound imaging was evaluated on Difference and agreement in common carotid artery intima-media thickness (cIMT) measurements between B-mode and M-mode ultrasound (MD 0.03 mm, 95% CI 0.02 - 0.05, p=<0.001). B-mode and M-mode ultrasound differed in measuring cIMT with poor inter-mode agreement and higher reproducibility for B-mode.