Key result
Intra-atrial ECG guidance achieves ~95% success in detecting catheter position during external jugular cannulation.
Why the study?
The reliability of using an intra-atrial electrocardiogram to guide external jugular vein catheterization required investigation.
Does intra-atrial electrocardiogram guidance reliably detect proper catheter tip placement and malposition during right external jugular vein catheterization in patients undergoing open heart surgery?
Observational (n=50)
Radiologist determining catheter position on chest X-ray was blinded to P-wave changes
prospective, randomized, single-centre study (single-arm evaluation)
No
Does intra-atrial electrocardiogram guidance reliably detect proper catheter tip placement and malposition during right external jugular vein catheterization in patients undergoing open heart surgery?
Intra-atrial ECG guidance reliably detects catheter tip malposition (95% accuracy) during external jugular vein cannulation, making the right EJV a safe alternative for central venous access in open heart surgery when the internal jugular vein is unavailable.
Answers are grounded in this paper and cite their sources.
Intra-atrial ECG may aid malposition detection in right EJV cannulation; hypothesis-generating and requires prospective validation.
Purpose: To investigate the reliability of intra-atrial electrocardiogram (ECG) use for external jugular vein (EJV) catheterization. Materials and Methods: Patients undergoing open heart surgery in Suleyman Demirel University Hospital between February and June 2006 were included in the study. Using a sterile Seldinger technique, a triple lumen polyurethane central venous catheter was introduced (Certofix® Trio V 720, length 20 cm, 7 French) under intra-atrial ECG guidance. The presence of an increase in P-wave size was recorded. Just after the surgery, a portable chest X-ray was taken. The method was considered to be successful when a change in P-wave could be seen and the catheter was in the superior vena cava, as well as when there was no change in P-wave and the catheter was not in the superior vena cava. Results: In six patients (12%), we were not able to advance the guidewire. In the remaining 44 patients, the catheter was inserted without problem. Eight of these 44 catheters were positioned in the innominate vein, with a malposition ratio of 18%. The success rate of external jugular vein cannulation with intra-atrial ECG was 95%. No complications occured related to the EJV cannulation. Conclusion: Considering that it is easily accessed without complication, and the malposition is successfully detected by intra-atrial ECG, EJV is a suitable access for central venous cannulation when internal jugular vein (IJV) is not usable.
No takes yet. Share an insight, caveat, or question.
Karaaslan et al. (2009) conducted an observational in Patients undergoing open heart surgery requiring central venous catheterization (n=50). Right external jugular vein (EJV) central venous catheterization under intra-atrial electrocardiogram (ECG) guidance was evaluated on Success rate of external jugular vein cannulation with intra-atrial ECG (defined as presence of P-wave increase when catheter is in the superior vena cava, or absence of P-wave increase when catheter is not in the superior vena cava). Intra-atrial ECG-guided external jugular vein cannulation achieved a 95% success rate in detecting catheter position, with a technical insertion success rate of 88% and an 18% malposition rate.