We describe a dual-wavelength, non-dispersive, spectro-radiometer that is optimized to measure carbon dioxide. This CO2 sensor was designed for use with a portable system for measuring water loss and photosynthesis in plant leaves. The elements of the sensor include: a highly insulated, miniature, blackbody source of infrared radiation; an external, folded, optical-absorption cell; alternating, narrow-band pass filters; and a commercial pyroelectric detector. Detector signals are filtered, amplified, and fed to a processing system based on a voltage-to-frequency converter that computes a ratio of the absorbed wavelength to a non-absorbed wavelength. The system, minus battery, weighs less than 1.5 kg and its optical head volume is 800 cm. The sensor requires 10 W of power from a 12-V source, and has an accuracy of about ±0.5 ppm at ambient (330 μ1/1) concentrations.
No takes yet. Share an insight, caveat, or question.
Bingham et al. (1981) studied this question.