Answers are grounded in this paper and cite their sources.
Hydrogen peroxide (H2O2) detection was demonstrated with multi-pass absorption spectroscopy using a commercial 76-m astigmatic multi-pass absorption cell. An ∼7.73-μm continuous wave, distributed feedback quantum cascade laser (CW DFB-QCL) was employed for targeting a strong H2O2 line (1296.2 cm-1) in the fundamental absorption band. Wavelength modulation spectroscopy combined with a second harmonic detection technique was utilized to increase the signal-to-noise ratio. By optimizing the pressure inside the multi-pass cell and the wavelength modulation depth, a minimum detection limit (1σ) of 13.4 ppbv was achieved for H2O2 with a 2-s sampling time. From an Allan-Werle deviation plot, the detection limit could be improved to 1.5 ppbv with an averaging time of 200 s. Interference effects of atmospheric air components are also discussed.
No takes yet. Share an insight, caveat, or question.
Cao et al. (2014) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: