Key result
In vivo wear of Pyrolytic carbon valve occluders from 35 Bjork-Shiley prostheses differed from in vitro results, with Hertzian impact wear theory estimating in vivo dynamic loads of 1 to 22 Newtons.
Hertzian impact wear theory can estimate in vivo dynamic loads on prosthetic heart valve occluders, explaining differences between in vivo and in vitro wear.
Answers are grounded in this paper and cite their sources.
May caution against extrapolating in vitro wear data to pyrolytic carbon valves; leaves open validation of Hertzian theory for in vivo load estimation.
Pyrolytic carbon (Pyrolite® carbon) valve occluders from 18 clinical Bjork‐Shiley prostheses recovered at necropsy were examined for wear. In situ durations extended to 12 years. Results were combined with data from 17 recovered valves in the literature to produce a data base of 35 Bjork‐Shiley valves. Cumulative in vivo results were compared to in vitro, Shiley type, accelerated life‐cycle tester results for Bjork‐Shiley prostheses. Comparative results were then interpreted using Hertzian impact wear theory. The theory provided a rational basis for the difference in wear observed in terms of applied dynamic load levels. Furthermore, Hertzian impact wear theory yielded an estimate of in vivo total dynamic loads on the order of 1 to 22 Newtons (0.2–5 lbf.) acting on the occluder of the heart valve prosthesis.
No takes yet. Share an insight, caveat, or question.
More et al. (1990) studied Prosthetic heart valve wear (n=35). In vivo clinical use vs. In vitro accelerated life-cycle testing was evaluated on Valve occluder wear and estimated dynamic loads. In vivo wear of Pyrolytic carbon valve occluders from 35 Bjork-Shiley prostheses differed from in vitro results, with Hertzian impact wear theory estimating in vivo dynamic loads of 1 to 22 Newtons.
Synapse has enriched 10 related papers. Consider them for comparative context: