Key points are not available for this paper at this time.
Answers are grounded in this paper and cite their sources.
Electrochemical weld corrosion explains Björk-Shiley strut fractures; leaves open targeted surveillance needs in exposed patients.
Welded downstream struts of Björk-Shiley Convexo-Concave heart valves show failure in vivo, but not in in vitro testing. A pyrolytic carbon pivoting disk occluder closes against a Haynes 25 alloy ring, which is electrochemically machined from solid with the upstream retaining struts. The weld area is de-alloyed, with residual porosity and carbide inclusions. The valve becomes a short-circuited electrochemical cell when fully open or closed. It is an aggressive chloride electrolyte, whose high pulsed flow (2 m/s) ensures that supply of oxygen-rich cathode reactant is not mass-transport-limited. During the flight of the occluder, the cell is randomly at open circuit. A random current pulse is applied to the metal parts on circuit closure. Failure is not from simple mechanical fatigue, but from stress-corrosion-cracking and erosion of the less noble weld area caused by these pulses. All welded valves of this type may be susceptible to ultimate in vivo failure.
No takes yet. Share an insight, caveat, or question.
Xiao et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: