Key result
No differences in embolic activity were observed when either 100% nitrogen or 95% oxygen-5% carbon dioxide gases were swept through a bubble oxygenator in an in vitro circuit.
Why the study?
Does sweeping 100% nitrogen gas reduce gaseous microemboli compared to a 95% O2-5% CO2 mixture in a bubble oxygenator?
Does sweeping 100% nitrogen gas reduce gaseous microemboli compared to a 95% O2-5% CO2 mixture in a bubble oxygenator?
Nitrogen gas does not reduce microembolic activity in bubble oxygenators compared to carbogen, and given the greater harm of nitrogen emboli, its use should be contraindicated.
No takes yet. Share an insight, caveat, or question.
No embolic benefit of nitrogen sweep versus carbogen seen in vitro; leaves open clinical relevance and safety of nitrogen in bubble oxygenators.
J. Mandl (1977) studied Cardiopulmonary bypass. 100% Nitrogen gas vs. 95% O2-5% CO2 mixture was evaluated on Embolic activity detected by ultrasonic Doppler. No differences in embolic activity were observed when either 100% nitrogen or 95% oxygen-5% carbon dioxide gases were swept through a bubble oxygenator in an in vitro circuit.