Key result
No differences in embolic activity were observed when either 100% nitrogen or 95% oxygen-5% carbon dioxide gases were swept through a bubble oxygenator in an in vitro circuit.
Why the study?
Does sweeping 100% nitrogen gas reduce gaseous microemboli compared to a 95% O2-5% CO2 mixture in a bubble oxygenator?
Does sweeping 100% nitrogen gas reduce gaseous microemboli compared to a 95% O2-5% CO2 mixture in a bubble oxygenator?
Nitrogen gas does not reduce microembolic activity in bubble oxygenators compared to carbogen, and given the greater harm of nitrogen emboli, its use should be contraindicated.
Answers are grounded in this paper and cite their sources.
No embolic benefit of nitrogen sweep versus carbogen seen in vitro; leaves open clinical relevance and safety of nitrogen in bubble oxygenators.
An ultrasonic Doppler was used to monitor embolic activity in the arterial line of an extracorporeal circuit. No differences in embolic activity were observed when either 100% N2 or 95% O2-5% CO2 gases were swept through a bubble oxygenator. Information is presented suggesting that nitrogen or air micro emboli are more deleterious than oxygen or carbogen microemboli. Consequently, this article recommends that the use of compressed air or nitrogen should be contraindicated in bubble oxygenators. Since the cardiotomy suction is a major source of nitrogen microemboli, this article also emphasizes that the cardiotomy suction be minimized and that a pressurized cardiotomy return be prevented from directly entering the oxygenator.
No takes yet. Share an insight, caveat, or question.
J. Mandl (1977) studied Cardiopulmonary bypass. 100% Nitrogen gas vs. 95% O2-5% CO2 mixture was evaluated on Embolic activity detected by ultrasonic Doppler. No differences in embolic activity were observed when either 100% nitrogen or 95% oxygen-5% carbon dioxide gases were swept through a bubble oxygenator in an in vitro circuit.