Key result
Day-to-day reproducibility of Holter repolarisation analysis was acceptable, with the best reproducibility observed for QT measured in narrow heart rate windows (coefficient of reproducibility 12ms).
Why the study?
Does Holter beat-by-beat analysis provide reproducible day-to-day measurements of repolarisation parameters in patients with and without cardiac disease?
Observational (n=30)
Does Holter beat-by-beat analysis provide reproducible day-to-day measurements of repolarisation parameters in patients with and without cardiac disease?
Day-to-day reproducibility of Holter repolarisation analysis is acceptable, with QT measurement in narrow heart rate windows demonstrating the best reproducibility.
Answers are grounded in this paper and cite their sources.
Supports narrow-window QT analysis for reproducible Holter repolarisation research; leaves clinical adoption open pending prospective validation.
UNLABELLED: Reproducibility of Holter QT analysis is not well established and has been assessed only in one study. STUDY DESIGN: We evaluated the day-to-day reproducibility of different QT parameters--mean and max (four beats basis) 24h QT and QTc (Bazett formula), QT for heart rate 55-60 [QT60], 75-80 [QT80] and 95-100 [QT100] beats/min and QT/RR slope (calculated in moving window of 3000 beats in 50 beat steps). QT intervals were measured from 48h digital ECG (sampled at 256 Hz) recordings using Del Mar Medical's QT software in beat-to-beat fashion. The analysed group consisted of 6 women and 24 men--13 patients with hypertrophic cardiomyopathy, 5 healthy family members of the patients with hypertrophic cardiomyopathy, 7 patients with CAD and 5 with other diseases (hypertension, arrhythmia, aborted sudden death without organic heart disease). Reproducibility was analysed with the methods proposed by Bland and Altman. RESULTS: The overall reproducibility of repolarisation parameters was acceptable. Coefficient of reproducibility for mean 24h QT was 24ms, mean QTc 12ms, max QT 22ms, max QTc 24ms. The best reproducibility was observed for QT60, QT80 and QT100 - 12ms, respectively. The poorest day-to-day reproducibility was recorded for the QT/RR slope parameters, which was related to lower heart rate reproducibility. CONCLUSIONS: We can conclude that day-to-day reproducibility of Holter repolarisation analysis is acceptable. QT measurement in narrow heart rate windows has the best reproducibility. Accurate QT analysis requires good quality recording, T wave amplitude above 0.2mV and an interactive QT measurement tool which includes verification, editing abilities.
No takes yet. Share an insight, caveat, or question.
Baranowski et al. (2003) conducted an observational in Hypertrophic cardiomyopathy, coronary artery disease, and other conditions (n=30). Holter beat-by-beat analysis of repolarisation (QT parameters) was evaluated on Day-to-day reproducibility of QT parameters. Day-to-day reproducibility of Holter repolarisation analysis was acceptable, with the best reproducibility observed for QT measured in narrow heart rate windows (coefficient of reproducibility 12ms).