Key result
The OptiBP smartphone application demonstrated good agreement with invasive arterial monitoring for mean arterial pressure with a bias of 0.9 mmHg, fulfilling AAMI/ISO universal standards.
Why the study?
Measuring blood pressure with a smartphone application could provide more widespread access to hypertension screening and monitoring compared to standard oscillometric methods.
Does the OptiBP optical smartphone application accurately measure blood pressure compared to invasive arterial monitoring in ICU patients?
Observational (n=30)
No
Does the OptiBP optical smartphone application accurately measure blood pressure compared to invasive arterial monitoring in ICU patients?
The OptiBP smartphone application showed good agreement with invasive arterial monitoring for mean and diastolic blood pressure in ICU patients, though systolic pressure measurements did not meet AAMI/ISO standards.
Answers are grounded in this paper and cite their sources.
Supports non-invasive BP options in ICU; leaves open need for randomized validation before clinical adoption.
BackgroundArterial hypertension is a public health problem worldwide. It is currently diagnosed and monitored non-invasively using the oscillometric method. Being able to measure blood pressure (BP) using a smartphone application could provide more widespread access to hypertension screening and monitoring. In this observational study in intensive care unit patients, we compared blood pressure values obtained using a new optical smartphone application (OptiBP™; test method) with arterial BP values obtained using a radial artery catheter (reference method).MethodsWe simultaneously measured three BP values every hour for five consecutive hours on two consecutive days using the two methods. Bland-Altman and error grid analyses were used for agreement analysis between both methods. The performance of the smarphone application was investigated using the Association for the Advancement of Medical Instrumentation (AAMI) and the International Organization for Standardization (ISO) definitions, which require the bias ± SD between two technologies to be below 5 ± 8 mmHg.ResultsAmong the 30 recruited patients, 22 patients had adequate OptiBP™ values and were thus analyzed. The Bland-Altman analysis revealed a mean of the differences ± SD between both methods of 0.9 ± 7 mmHg for mean arterial pressure (MAP), 0.2 ± 14 mmHg for systolic arterial pressure (SAP), and 1.1 ± 6 mmHg for diastolic arterial pressure (DAP). Error grid analysis demonstrated that the proportions of measurement pairs in risk zones A to E were 88.8% (no risk), 10% (low risk), 1% (moderate risk), 0% (significant risk), and 0% (dangerous risk) for MAP and 88.4%, 8.6%, 3%, 0%, 0%, respectively, for SAP.ConclusionsThis method comparison study revealed good agreement between BP values obtained using the OptiBP™ and those done invasively. The OptiBP™ fulfills the AAMI/ISO universal standards for MAP and DAP (but not SAP). Error grid showed that the most measurements (≥ 97%) were in risk zones A and B.ClinicalTrials.gov registrationNCT04728477
No takes yet. Share an insight, caveat, or question.
Desebbe et al. (2022) conducted an observational in Intensive Care / Arterial Blood Pressure Monitoring (n=30). OptiBP smartphone application vs. Invasive arterial blood pressure monitoring (radial artery catheter) was evaluated on Agreement (bias) between OptiBP and invasive arterial blood pressure for mean arterial pressure (MAP) (MD 0.9 mmHg, 95% CI -13 to 15). The OptiBP smartphone application demonstrated good agreement with invasive arterial monitoring for mean arterial pressure with a bias of 0.9 mmHg, fulfilling AAMI/ISO universal standards.