Key result
The OptiBP smartphone application fulfilled AAMI/ESH/ISO universal validation standards, demonstrating a mean difference of 0.5 ± 7.7 mmHg for SBP and 0.4 ± 4.6 mmHg for DBP compared to auscultatory reference.
Why the study?
The study was designed to assess the accuracy of the OptiBP smartphone application in estimating blood pressure according to AAMI/ESH/ISO universal standards.
Does the OptiBP smartphone application accurately estimate blood pressure compared to standard auscultatory measurement in a general adult population?
Observational (n=91)
Blinded observers
No
Does the OptiBP smartphone application accurately estimate blood pressure compared to standard auscultatory measurement in a general adult population?
Answers are grounded in this paper and cite their sources.
Supports OptiBP feasibility for cuffless estimation; leaves open larger validation and outcome studies before clinical use.
Objective The aim of this study was to assess the accuracy of the OptiBP mobile application based on an optical signal recorded by placing the patient’s fingertip on a smartphone’s camera to estimate blood pressure (BP). Measurements were carried out in a general population according to existing standards of the Association for the Advancement of Medical Instrumentation (AAMI), the European Society of Hypertension (ESH) and the International Organization for Standardization (ISO). Methods Participants were recruited during a scheduled appointment at the hypertension clinic of Lausanne University Hospital in Switzerland. Age, gender and BP distribution were collected to fulfill AAMI/ESH/ISO universal standards. Both auscultatory BP references and OptiBP were measured and compared using the opposite arm simultaneous method as described in the 81060-2:2018 ISO norm. Results A total of 353 paired recordings from 91 subjects were analyzed. For validation criterion 1, the mean ± SD between OptiBP and reference BP recordings was respectively 0.5 ± 7.7 mmHg and 0.4 ± 4.6 mmHg for SBP and DBP. For validation criterion 2, the SD of the averaged BP differences between OptiBP and reference BP per subject was 6.3 mmHg and 3.5 mmHg for SBP and DBP. OptiBP acceptance rate was 85%. Conclusion The smartphone embedded OptiBP cuffless mobile application fulfills the validation requirements of AAMI/ESH/ISO universal standards in a general population for the measurement of SBP and DBP.
No takes yet. Share an insight, caveat, or question.
Degott et al. (2021) conducted an observational in Hypertension / General Population (n=91). OptiBP mobile application vs. Auscultatory blood pressure reference was evaluated on Accuracy of OptiBP compared to reference BP (AAMI/ESH/ISO validation criterion 1) (MD 0.5 ± 7.7 mmHg (SBP) and 0.4 ± 4.6 mmHg (DBP)). The OptiBP smartphone application fulfilled AAMI/ESH/ISO universal validation standards, demonstrating a mean difference of 0.5 ± 7.7 mmHg for SBP and 0.4 ± 4.6 mmHg for DBP compared to auscultatory reference.