Key result
The OptiBP smartphone application fulfilled AAMI/ESH/ISO universal validation standards, demonstrating a mean difference of 0.5 ± 7.7 mmHg for SBP and 0.4 ± 4.6 mmHg for DBP compared to auscultatory reference.
Why the study?
The study was designed to assess the accuracy of the OptiBP smartphone application in estimating blood pressure according to AAMI/ESH/ISO universal standards.
Does the OptiBP smartphone application accurately estimate blood pressure compared to standard auscultatory measurement in a general adult population?
Population
91 subjects recruited at a hypertension clinic in Switzerland
Comparison
OptiBP smartphone application vs reference auscultatory blood pressure measurement
Design
Validation study using the opposite arm simultaneous method
Authors
Loading...
Supports OptiBP feasibility for cuffless estimation; leaves open larger validation and outcome studies before clinical use.
Observational (n=91)
Blinded observers
No
Does the OptiBP smartphone application accurately estimate blood pressure compared to standard auscultatory measurement in a general adult population?
Mean Difference: 0.5
Degott et al. (2021) conducted an observational in Hypertension / General Population (n=91). OptiBP mobile application vs. Auscultatory blood pressure reference was evaluated on Accuracy of OptiBP compared to reference BP (AAMI/ESH/ISO validation criterion 1) (MD 0.5 ± 7.7 mmHg (SBP) and 0.4 ± 4.6 mmHg (DBP)). The OptiBP smartphone application fulfilled AAMI/ESH/ISO universal validation standards, demonstrating a mean difference of 0.5 ± 7.7 mmHg for SBP and 0.4 ± 4.6 mmHg for DBP compared to auscultatory reference.