Key result
An end-to-end standard-based patient monitoring solution using ISO/IEEE 11073 and EN13606 was successfully implemented and tested in a laboratory environment to demonstrate feasibility.
Demonstrates the technical feasibility of an end-to-end standard-based patient monitoring solution for ICUs using ISO/IEEE 11073 and EN13606.
This is the authors' abstract. We don't add key points for this paper.
A proof-of-concept design of a patient monitoring solution for intensive care unit environments has been presented. It is end-to-end standard-based, using ISO/IEEE 11073 (X73) in the bedside environment and EN13606 to communicate the information to an electronic healthcare record (EHR) server. At the bedside end, the system is a plug-and-play sensor network communicating with a gateway that collects medical information and sends the data to a monitoring server. The monitoring server transforms this information into an EN13606 extract to be stored on the EHR server. The system has been implemented to comply with the last X73 and EN13606 available versions and tested in a laboratory environment to demonstrate the feasibility of an end-to-end standard-based solution.
Answers are grounded in this paper and cite their sources.
Laboratory feasibility of standardized ICU monitoring encourages further development; leaves open real-world clinical efficacy and workflow integration.
No takes yet. Share an insight, caveat, or question.
Fernández et al. (2008) studied Intensive care unit monitoring. End-to-end standard-based patient monitoring solution (ISO/IEEE 11073 and EN13606) was evaluated. An end-to-end standard-based patient monitoring solution using ISO/IEEE 11073 and EN13606 was successfully implemented and tested in a laboratory environment to demonstrate feasibility.