Why the study?
Does a phosphorylcholine-coated, closed cardiopulmonary bypass system improve safety and efficacy compared to conventional CPB and OPCAB in patients undergoing first-time CABG?
Cohort (n=88)
Does a phosphorylcholine-coated, closed cardiopulmonary bypass system improve safety and efficacy compared to conventional CPB and OPCAB in patients undergoing first-time CABG?
Answers are grounded in this paper and cite their sources.
The ECC.O closed cardiopulmonary bypass system is safe and efficacious, significantly reducing hemodilution, transfusion requirements, and inflammatory response compared to conventional CPB.
We designed a pilot study to assess as primary end point the safety and efficacy of a new phosphorylcholine-coated, closed cardiopulmonary bypass (CPB) system (extracorporeal circulation, optimized [ECC.O], Dideco, Mirandola, Italy). The secondary end point was to compare results with two retrospectively matched cohorts of patients who underwent isolated coronary artery by-pass graft (CAGB) with nonphosphorylcholine-bonded circuits and cardiotomy suction (Group II, n = 32) and off-pump coronary artery by-pass (OPCAB) (Group III, n = 26). In January 2005, 30 patients (Group I) undergoing first-time CABG were assigned to the ECC.O group. Five minutes after CPB, initial hematocrit levels were significantly and consistently highest in Group I relative to Group II (Group I, 29.7 +/- 4.4 vs. Group II, 22.7 +/- 4.1; P < 0.001). Red blood cell transfusion rate was reduced drastically in Group I versus Group II (P < 0.001). High differences were also observed in C-reactive protein levels at 24 h after surgery (Group I vs. Group II-P < 0.001 and vs. Group III-P < 0.001) and at 72-h peak value (Group I vs. Group II-P < 0.001 and vs. Group III-P < 0.001). The routine clinical use of the ECC.O system has been demonstrated to be both clinically safe and efficacious. An intensive training program for surgeons, perfusionists, and anesthesiologists is required.
No takes yet. Share an insight, caveat, or question.
Agati et al. (2007) conducted a cohort in First-time coronary artery bypass graft (CABG) (n=88). Phosphorylcholine-coated, closed cardiopulmonary bypass (CPB) system (ECC.O) vs. Nonphosphorylcholine-bonded circuits with cardiotomy suction (Group II) and off-pump CABG (Group III) was evaluated on Safety and efficacy (assessed via hematocrit levels 5 minutes after CPB) (p=<0.001).
Synapse has enriched 10 related papers. Consider them for comparative context: