Latex immunoassay is a nonisotopic method based on agglutination, by protein, of calibrated latex particles coated with a specific antibody. The assay has been automated in a simple continuous-flow system by incubating the reaction mixture in a heated mixing coil for 25 min and measuring the agglutination with a cell counter. No external shaking of the latex suspension and no additional reagent is required for the agglutination. The method can accurately and precisely quantify a wide variety of proteins in plasma and urine, including human ferritin, beta 2-microglobulin, retinol-binding protein, and albumin. Depending on the antigen-antibody system, the detection limit ranges from 10(-10) to about 10(-12) mol/L. Within- and between-assay CVs are less than 10%. In the assay of ferritin, sera are pretreated to eliminate interferences from chylomicrons, complement, and rheumatoid factor.
No takes yet. Share an insight, caveat, or question.
Bernard et al. (1983) studied this question.