Key result
The Cook Pigtail central venous catheter had a lower median ease of insertion rating compared to standard CVCs and caused more arrhythmias during insertion.
Why the study?
Does the Cook Pigtail central venous catheter improve ease of insertion and safety compared to standard central venous catheters in adults undergoing elective surgery?
RCT (n=203)
Does the Cook Pigtail central venous catheter improve ease of insertion and safety compared to standard central venous catheters in adults undergoing elective surgery?
The Cook Pigtail CVC was associated with lower ease of insertion and a higher incidence of insertion-related arrhythmias compared to standard CVCs, though it may be preferred for long-term use.
Answers are grounded in this paper and cite their sources.
Cook Pigtail CVCs reduce ease of insertion and raise arrhythmia rates; challenges preference for pigtail design over standard CVCs in elective surgery.
The Cook Pigtail central venous catheter (CVC) has been designed to diminish the risk of vascular perforation and consequent cardiac tamponade. With the participation of 12 consultant anaesthetists and 19 registrars, adults undergoing elective surgery were randomized to receive either a Pigtail (n = 101) or their consultant anaesthetists' "standard" CVC (n = 102). Median ease of insertion was rated 8 for Pigtail CVCs and 9 for standards (10 being best; P = 0.001). Arrhythmias occurred during 16 standard and 33 Pigtail central venous catheter insertions (P < 0.006). No significant difference was found in insertion time or radiographically assessed tip depth for standard and Pigtail central venous catheters. A perforated right atrium of uncertain cause occurred in a patient who received an Arrow triple-lumen central venous catheter. Participating consultant anaesthetists preferred their "standard" central venous catheter for routine use, but five indicated that they would select a Cook Pigtail where long-term use was planned because of in vitro evidence of its greater safety.
No takes yet. Share an insight, caveat, or question.
Merry et al. (1999) conducted an RCT in Elective surgery (n=203). Cook Pigtail central venous catheter vs. Standard central venous catheter was evaluated on Median ease of insertion (rated 1-10, 10 being best) (p=0.001). The Cook Pigtail central venous catheter had a lower median ease of insertion rating compared to standard CVCs and caused more arrhythmias during insertion.
Synapse has enriched 10 related papers. Consider them for comparative context: