Why the study?
The performance of the TEG 6s compared with the TEG 5000 had not been reported in patients on extracorporeal membrane oxygenation.
Does the TEG 6s analyzer provide comparable results to the TEG 5000 analyzer in hospitalized adult patients on ECMO?
Observational (n=34)
No
Does the TEG 6s analyzer provide comparable results to the TEG 5000 analyzer in hospitalized adult patients on ECMO?
The TEG 6s device shows acceptable correlation with the TEG 5000 in ECMO patients, but wide limits of agreement suggest they may not be interchangeable outside reference ranges.
Answers are grounded in this paper and cite their sources.
TEG 6s may serve as surrogate for TEG 5000 on ECMO but wide limits of agreement advise caution outside reference ranges; leaves open interchangeability.
Thromboelastography (TEG) evaluates viscoelastic properties of blood clot formation. The TEG 5000 analyzer is commonly used but prone to errors related to vibration or operator error. The TEG 6s was developed to overcome these limitations. Performance of TEG 6s compared with TEG 5000 has not been reported in extracorporeal membrane oxygenation (ECMO). We compared the agreement between devices via a single-center prospective observational study in hospitalized adult patients on ECMO. Data for both devices were collected daily for 3 days after ECMO initiation. Standard tests for method comparison were used. Thirty-four matching samples were available for analysis. Minimal bias was noted; however, the limit of agreement was wide for TEG parameters. Visually, agreement was better for values within the reference ranges of the tests. Lin's coefficients demonstrated moderate correlation for reaction time and alpha angle (0.58; 95% confidence interval [CI], 0.31-0.76 and 0.63; 95% CI, 0.40-0.78, respectively). Excellent correlation was demonstrated for kinetic time and maximum amplitude (0.88; 95% CI, 0.79-0.94 and 0.89; 95% CI, 0.79-0.94). The TEG 6s device may represent an acceptable surrogate for the TEG 5000 in patients on ECMO. However, limitations in reliability were noted, and the devices may not be interchangeable when results fall outside of the reference values.
No takes yet. Share an insight, caveat, or question.
Chandel et al. (2021) conducted an observational in Extracorporeal membrane oxygenation (ECMO) (n=34). TEG 6s analyzer vs. TEG 5000 analyzer was evaluated on Agreement between devices for maximum amplitude (Lin's coefficient 0.89, 95% CI 0.79-0.94).