We report the results of a national quality-control survey on glycohemoglobin (GHb), monitored in France by the Société Française de Biologie Clinique on behalf of the authority of the "Agence du Médicament." A sample of lyophilized hemolysate was sent to 3109 laboratories. Results were obtained from 2770 laboratories. HbA1C, HbA1, and total GHb were measured by 50%, 24%, and 26% of the participants, respectively. Of these measurements, 79% of the HbA1C results and 76% of the total GHb results, but only 48% of the HbA1 results, were within the +/- 20% limits of the indicated target values. Mean values for the hemolysate ranged from 8% to 11% for HbA1C, from 7% to 12% for HbA1, and from 11% to 13% for total GHb. The interlaboratory CVs ranged from 3% to 20%, according to method used. So, methods used for GHb assay, which are based on various principles, exhibit very different analytical performances. Nonetheless, this large-scale study indicates that some techniques can support transferability of results from laboratory to laboratory.
No takes yet. Share an insight, caveat, or question.
Gillery et al. (1995) studied this question.