Key result
The Elecsys Periostin immunoassay demonstrated acceptable analytical performance, including a limit of quantitation of 10 ng/mL and reproducibility CVs of 1.7-3.1% across 3 testing sites.
The Elecsys Periostin immunoassay demonstrates robust analytical performance and reproducibility, making it suitable for measuring serum periostin at clinically relevant concentrations.
Answers are grounded in this paper and cite their sources.
Facilitates standardized periostin measurement in IL-13 asthma research; leaves open clinical utility pending outcome studies.
OBJECTIVE: The multifunctional cytokine IL-13 is thought to play a central role in Type 2 inflammation in asthma. Serum periostin has been explored as a candidate biomarker for evaluating IL-13 activity in the airway. We describe the technical performance characteristics of a novel, fully automated immunoassay for the determination of periostin in serum. DESIGN AND METHODS: Limit of blank [LoB], limit of detection [LoD] and limit of quantitation [LoQ], linearity, precision and reproducibility across sites and lots were evaluated according to Clinical and Laboratory Standards Institute guidelines. Interferences and sample stability were also investigated. RESULTS: The pre-specified values for LoB (2ng/mL), LoD (4ng/mL) and LoQ (10ng/mL) were met. The assay was linear throughout the measuring range (10-160ng/mL) with recoveries within ±10% of target at concentrations >30ng/mL and within ±3ng/mL at concentrations ≤30ng/mL. Recovered periostin concentrations were also within ±10% of target in presence of 43 potentially interfering substances and drugs. Samples were stable across various storage conditions and durations (24h at room temperature, 7days at 4°C, 12weeks at -20°C, and 3 freeze/thaw cycles). Repeatability experiments resulted in CVs across samples and controls ranging from 0.9-1.5%. Intermediate precision was 1.2-1.7% and reproducibility including 3 testing sites and 3 reagent lots was 1.7-3.1%. The final assay correlates to the assay version used in previous clinical trials (Pearson's r=0.998, bias at 50ng/mL=1.2%). CONCLUSION: Periostin immunoassay including a multicenter precision analysis demonstrated that the assay is suitable for measuring serum periostin at clinically important concentrations around 50ng/mL.
No takes yet. Share an insight, caveat, or question.
Palme et al. (2016) studied Asthma. Elecsys Periostin immunoassay was evaluated on Limit of blank, limit of detection, limit of quantitation, linearity, precision and reproducibility. The Elecsys Periostin immunoassay demonstrated acceptable analytical performance, including a limit of quantitation of 10 ng/mL and reproducibility CVs of 1.7-3.1% across 3 testing sites.
Synapse has enriched 10 related papers. Consider them for comparative context: