A new spectrometer for the precision measurement of dielectric permittivity and loss tangent is presented. The new instrument is capable of providing high-resolution data for the first time over an extended W-band (68-118 GHz) frequency for specimens with a large range of absorption values, including highly absorbing specimens that otherwise would not be possible. A novel technique based on the unbalanced bridge is developed for the measurement of the phase of the wave passed through the specimen in free space (quasi-optical) with reference provided by a waveguide arm. Specially constructed precision waveguide and quasi-optical components allowed reliable broadband operation. A number of common dielectrics are measured, and results are compared with previously reported data.