Key result
Extreme stent malapposition occurred in 19.2% of bifurcation lesions treated with kissing balloon inflation, associated with a lower stent eccentricity index.
Why the study?
Anecdotal cases indicated that incorrect kissing balloon inflation can cause extreme stent malapposition or crushing, but the occurrence of this finding despite acceptable angiographic results was unquantified.
Observational (n=229)
Nearly 20% of bifurcation lesions treated with kissing balloon inflation exhibit extreme stent malapposition on OCT, likely due to incorrect guidewire passage during exchange.
Answers are grounded in this paper and cite their sources.
Stent malapposition after kissing balloon inflation warrants procedural caution in bifurcations; cross-sectional data leaves open whether eccentricity index affects outcomes.
Background: Anecdotal cases of bifurcation lesion stenting indicated that a wrong kissing Balloon Inflation (KBI) technique can lead to extreme stent malapposition with even stent crushing. The aim of this study was to quantify the occurrence of this finding, despite achieving acceptable angiographic results. Methods: A total of 229 bifurcation lesions were included in this study. Optical Coherence Tomography (OCT) was obtained immediately after KBI. The stented bifurcation lesions were classified into 2 groups: group 1 with extreme malapposition (malapposition distance ≥ 1mm and in-stent minimal Cross Sectional Area (CSA) <70% of the distal reference CSA), and group 2 (malapposition distance <1mm and/or in-stent CSA >70% of the distal reference CSA). Results: OCT revealed a mean stent malapposition distance of 0.43 ± 0.47 mm extending for a mean length of 1.88 ± 1.82 mm. Forty four cases (19.2%) of extreme stent malapposition were observed, and characterized by a stent malapposition distance of 1.49 ± 0.3 mm. These stents were under-expanded with lower stent eccentricity index (SEI), compared with those without extreme malapposition (48.38 ± 9.67 vs. 60.97 ± 33.29, p=0.46 and 0.57 ± 0.15 vs. 0.74 ± 0.89, p<0.001, respectively). This is mainly attributed to the wrong passage of guide-wires, while performing wire exchange during KBI technique. Conclusion: The bifurcation stenting requiring a KBI technique can be complicated by an extreme degree of stent malapposition; or even stent crushing in some cases, caused by the wrong passage of one or both guide-wires through the stents, during the wire exchange.
No takes yet. Share an insight, caveat, or question.
Hesham Refaat (2020) conducted an observational in Bifurcation lesions (n=229). Extreme stent malapposition vs. No extreme stent malapposition was evaluated on Occurrence of extreme stent malapposition. Extreme stent malapposition occurred in 19.2% of bifurcation lesions treated with kissing balloon inflation, associated with a lower stent eccentricity index.
Synapse has enriched 10 related papers. Consider them for comparative context: