This paper describes micromachined scanning thermocouple probes that exploit the low thermal conductivity and the high mechanical flexibility of polyimide as a structural material. They are surface micromachined using a low-temperature six-mask process suitable for appending to a CMOS fabrication sequence. The probes are 200-1000-/spl mu/m long, 40-120-/spl mu/m wide, and of varying thickness. They are assembled by a flip-over approach that eliminates the need for dissolving the substrate wafer or removing the probe from it. Temperature sensing is provided by thin-film Ni/W or chromel/alumel thermopiles embedded in the polyimide, which provide Seebeck coefficients of 22.5 and 37.5 /spl mu/V/K per junction, respectively. Modeling results indicate that the low thermal conductivity of polyimide causes the temperature drop along the probe length to be much higher than with other candidate materials such as Si or SiO/sub 2/, which contributes to improved thermal isolation of the sample and higher temperature sensitivity of the probe. However, the response time of the probe is compromised, and the measured -3 dB bandwidth of the probes is /spl ap/500 Hz. A sample scan is presented.
No takes yet. Share an insight, caveat, or question.
Li et al. (2001) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: