Answers are grounded in this paper and cite their sources.
GSAS instrument parameters are tabulated for a variety of laboratory and synchrotron diffractometers to give users an idea of the typical ranges of profile parameters when they generate their own instrument parameter files. For modern high-resolution laboratory diffractometers, the parameters fall in the ranges 0< U <3, V =0, 0< W <4, 1< X <3, 0< Y <3, 1 S / L <0.03. For synchrotron diffractometers, the parameters fall in the ranges 0< U <1.2, −1< V <0, 0< W <1, 0< X <1, 0< Y <1, 0 S / L <0.001, and 0< H / L <0.007. FULLPROF equivalents are also reported. The factors which are convoluted together to generate the instrument profile are described.
No takes yet. Share an insight, caveat, or question.
Kaduk et al. (2011) studied this question.