Key result
Atrial inhibited pacing in sick sinus node syndrome was associated with low rates of lead failure, atrial fibrillation, and atrioventricular block over a mean of 32 months.
Why the study?
What is the incidence of complications and the need for rate responsiveness in patients with sick sinus node syndrome receiving atrial inhibited pacemakers?
Observational (n=66)
What is the incidence of complications and the need for rate responsiveness in patients with sick sinus node syndrome receiving atrial inhibited pacemakers?
Atrial inhibited pacing is well suited for selected patients with sinus node dysfunction and preserved AV conduction, with a low incidence of complications and limited need for rate-responsive pacemakers.
Answers are grounded in this paper and cite their sources.
May support atrial inhibited pacing in selected sinus node dysfunction; leaves open need for randomized confirmation of durability.
A consecutive series of 66 patients (males = 32; mean age +/- SD = 71 +/- 9 years) given atrial inhibited pacemakers for sick sinus nodes were followed to study the incidence of lead failures, chronic atrial tachyarrhythmias, and atrioventricular conduction disturbances. The need for rate responsive pacing was also assessed. Pre and postoperative investigation could include carotid sinus massage, Holter monitoring, exercise testing, and invasive electrophysiology. The mean follow-up time +/- SD was 32 +/- 29 months (median = 26 months). Three patients (5%) had their pacemakers replaced due to lead failures (loss of sensing = 2; exit block = 1). Two pacemakers (3%) were replaced after 5 and 22 months due to atrial fibrillation. Four patients (6%) received new pacemakers because of development of second-degree or complete atrioventricular block after 1, 6, 12, and 31 months, respectively. During exercise, most patients (76%) responded with an increase in sinus rate at least as marked as that achievable with the currently available rate responsive pacemakers. Assuming careful patient selection, atrial inhibited pacing is well suited for many patients with sinus node dysfunction and preserved atrioventricular conduction. There is a limited need for rate responsive pacemakers in these patients.
No takes yet. Share an insight, caveat, or question.
Kallryd et al. (1989) conducted an observational in Sick sinus node syndrome (n=66). Atrial inhibited pacemakers was evaluated on Incidence of lead failures, chronic atrial tachyarrhythmias, and atrioventricular conduction disturbances. Atrial inhibited pacing in sick sinus node syndrome was associated with low rates of lead failure, atrial fibrillation, and atrioventricular block over a mean of 32 months.
Synapse has enriched 10 related papers. Consider them for comparative context: