Key result
Steerable sheath with a needle's eye snare safely extracts a chronically implanted Micra leadless pacemaker.
Why the study?
Data on leadless pacemaker extraction are limited to a handful of case reports, dedicated extraction tools are lacking, and use of a needle's eye snare to extract a leadless pacemaker had not been previously reported.
Can a needle's eye snare combined with a steerable sheath through a large-bore access sheath safely and effectively extract a chronically implanted leadless pacemaker in a patient with persistent Staphylococcus aureus bacteremia?
Case Report (n=1)
No
Can a needle's eye snare combined with a steerable sheath through a large-bore access sheath safely and effectively extract a chronically implanted leadless pacemaker in a patient with persistent Staphylococcus aureus bacteremia?
This case report is the first to demonstrate that a needle's eye snare combined with a steerable sheath through a large-bore introducer sheath can safely and effectively extract a chronically implanted Micra leadless pacemaker to resolve persistent bacteremia.
Answers are grounded in this paper and cite their sources.
Supports percutaneous extraction feasibility for infected leadless pacemakers; leaves open generalizability and safety in larger cohorts.
The Micra Transcatheter Pacing System (Medtronic, Minneapolis, MN) is a leadless pacemaker (LP) that is associated with a low rate of implant-related complications and, during initial experience, a very low infection rate by comparison with that observed for transvenous pacing systems.1 Potential indications for LP explantation/extraction include dislodgement, suboptimal pacing performance, the use of LPs as a bridge between permanent transvenous systems following device infection, or infection of the LP itself.
No takes yet. Share an insight, caveat, or question.
Mor-Welch et al. (2022) conducted a case report in Persistent Staphylococcus aureus bacteremia with a chronically implanted leadless pacemaker (n=1). Leadless pacemaker extraction using a 20 mm Needle's Eye Snare via an 8F steerable sheath and 23F Micra introducer sheath was evaluated on Successful extraction of the leadless pacemaker and resolution of persistent bacteremia. A needle's eye snare combined with a steerable sheath successfully extracted a 2-year-old Micra leadless pacemaker without complication, resolving persistent S. aureus bacteremia.
Synapse has enriched 10 related papers. Consider them for comparative context: