Key result
Insertion of a temporary parachute valve in a mechanical simulator increased diastolic pressure from 7 to 38 mm Hg, increased cardiac output from 2.08 to 4.66 L/min, and reduced regurgitant volume.
Why the study?
Does a wire-mounted, monoleaflet, temporary valve apparatus improve hemodynamic stability in a mechanical cardiovascular simulator?
Does a wire-mounted, monoleaflet, temporary valve apparatus improve hemodynamic stability in a mechanical cardiovascular simulator?
In an in vitro simulator, a temporary catheter-based aortic parachute valve improved hemodynamics, suggesting it may help maintain stability during off-pump aortic valve replacement.
Answers are grounded in this paper and cite their sources.
May support temporary valve development for off-pump procedures; leaves open translation from in vitro models to clinical use.
Recently developed technologies allow aortic valve implantation off-pump in a beating heart. In this procedure, the native, stenotic aortic valve is not removed, but simply crushed by a pressure balloon mounted on a percutaneous catheter. Removal of the native aortic cusps before valve replacement may reduce the incidence of annular or cuspal calcium embolization and late perivalvular leaks and increase implantable valve size. However, a temporary valve system in the ascending aorta may be necessary to maintain hemodynamic stability by reducing acute aortic regurgitation and left ventricular volume overload. This study evaluates the hemodynamic effects of a wire-mounted, monoleaflet, temporary valve apparatus in a mechanical cardiovascular simulator. Aortic flow, systemic pressure and left ventricular pressure were continuously monitored. An intraluminal camera obtained real-time proximal and distal images of the valve in operation. Insertion of the parachute valve in the simulator increased diastolic pressure from 7 to 38 mm Hg. Cardiac output increased from 2.08 to 4.66 L/min and regurgitant volume decreased from 65 to 23 mL. In conclusion, placement of a temporary valve in the ascending aorta may help maintain hemodynamic stability and improve off-pump aortic valve replacement.
No takes yet. Share an insight, caveat, or question.
Vandenberghe et al. (2008) studied Aortic regurgitation. Temporary catheter-based aortic parachute valve vs. Baseline (no valve) was evaluated on Hemodynamic effects (diastolic pressure, cardiac output, regurgitant volume). Insertion of a temporary parachute valve in a mechanical simulator increased diastolic pressure from 7 to 38 mm Hg, increased cardiac output from 2.08 to 4.66 L/min, and reduced regurgitant volume.
Synapse has enriched 10 related papers. Consider them for comparative context: