Key result
Cardiac output measurement using the Vigileo monitor was not clinically equivalent to pulmonary artery thermodilution, showing a bias of -0.1 l/min and a 48% percentage error compared to ICO.
Why the study?
Does cardiac output assessment using arterial pulse waveform analysis provide equivalent measurements to pulmonary arterial thermodilution in patients undergoing coronary artery bypass grafting?
Observational (n=9)
Does cardiac output assessment using arterial pulse waveform analysis provide equivalent measurements to pulmonary arterial thermodilution in patients undergoing coronary artery bypass grafting?
Mean Difference: -0.1
Arterial pulse waveform analysis using the Vigileo monitor is not clinically equivalent to continuous thermodilution by pulmonary artery catheter for measuring cardiac output during and after cardiac surgery.
Vigileo should not substitute for thermodilution cardiac output in CABG; leaves open need for device validation in perioperative cohorts.
Many devices are available to assess cardiac output (CO) in critically ill patients and in the operating room. Classical CO monitoring via a pulmonary artery catheter involves continuous cardiac output (CCO) measurement. The second generation of Flotrac/Vigileo monitors propose an analysis of peripheral arterial pulse waves to calculate CO (APCO) without calibration. The aim of our study was to compare the CO between the Swan Ganz catheter and the VigileoT. In this observational study, nine patients undergoing coronary artery bypass grafting were prospectively included. APCO, mean (CCO) and instantaneous CO (ICO) were measured. Perioperative and postoperative assessments were performed up to 24 hours post-surgery. Measurements were recorded every minute, resulting in the collection of 6492 data pairs. Comparison of APCO and ICO showed a limited bias of -0.1 l/min but an important percentage error of 48%. Corresponding values were -0.1 l/min and 46% for the APCO versus CCO comparison, and 0 and 17% for ICO versus CCO comparison. Large inter-individual variability does exist. During cardiac surgery and after leaving the operating room, Vigileo is not clinically equivalent to continuous thermodilution by pulmonary artery catheter Nevertheless, the connection between CCO and ICO relates the difference between APCO and CCO more to the different algorithms used. Further efforts should be concentrated on assessing the ability of this device to track changes in cardiac output.
No takes yet. Share an insight, caveat, or question.
Hamm et al. (2010) conducted an observational in Coronary artery bypass grafting (n=9). Vigileo monitor (APCO) vs. Swan Ganz catheter (pulmonary artery thermodilution) was evaluated on Comparison of cardiac output between APCO and instantaneous cardiac output (ICO) (bias -0.1 l/min, percentage error 48%). Cardiac output measurement using the Vigileo monitor was not clinically equivalent to pulmonary artery thermodilution, showing a bias of -0.1 l/min and a 48% percentage error compared to ICO.