This first reported case of calcific structural failure of a Toronto stentless porcine valve at 6 years raises potential concerns regarding its long-term durability.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to monitor Toronto stentless valves for early calcific failure; leaves open durability questions in larger cohorts.
We describe the calcific structural failure of a Toronto stentless porcine valve (TSPV) which had been used to replace a calcified bicuspid aortic valve in a 46-year-old man. Against expectations, left ventricular hypertrophy persisted and the transvalvular pressure gradient rose to 125 mmHg by 6 years with the patient becoming symptomatic and requiring redo surgery. On removal the TSPV showed atypical calcification of the leaflet hinges and wall. To our knowledge this is the first case reported and it may have implications for long term durability and future surgery using this prosthesis.
No takes yet. Share an insight, caveat, or question.
Tansley et al. (2000) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: