Key result
In a healthy ovine model, Mitra-Spacer implantation was feasible and safe up to 90 days, with 7 of 8 animals surviving and no evidence of structural damage or cardiac deterioration.
Why the study?
Does Mitra-Spacer implantation demonstrate safety and feasibility in a healthy ovine model?
Does Mitra-Spacer implantation demonstrate safety and feasibility in a healthy ovine model?
The Mitra-Spacer device demonstrated safety and feasibility in a healthy ovine model up to 90 days, supporting further development for mitral regurgitation treatment.
Answers are grounded in this paper and cite their sources.
Supports short-term safety in healthy sheep; leaves open testing in diseased models before human translation.
AIMS: The Mitra-Spacer (Cardiosolutions, Bridgewater, MA, USA) is designed to treat mitral regurgitation by introducing a dynamic spacer that constantly adapts to the changing haemodynamic conditions during the cardiac cycle. We aimed to evaluate the performance and safety of this device in the chronic ovine model. METHODS AND RESULTS: Eight sheep were enrolled in this study. Through a left thoracotomy, the Mitra-Spacer was inserted via the transapical approach and advanced into the left atrium (LA) under imaging guidance. Device performance and safety were evaluated up to 90 days using fluoroscopy, echocardiography and histopathology. The volume within the balloon spacer shifted during the cardiac cycle in all cases. Seven animals survived up to 90 days for terminal imaging and tissue harvest. Echocardiography showed no change in left ventricle (LV) ejection fraction from baseline to 90 days. There were no observations of changes in LV diastolic function, pulmonary vein inflow, or tricuspid valve function. Histological analysis demonstrated no significant injury to the mitral apparatus. CONCLUSIONS: In the healthy ovine model, Mitra-Spacer implantation was feasible and safe. At 90 days, no evidence of structural damage to the mitral apparatus or deterioration of cardiac performance was demonstrated.
No takes yet. Share an insight, caveat, or question.
Peppas et al. (2017) studied Healthy ovine model (device for mitral regurgitation) (n=8). Mitra-Spacer was evaluated on Device performance and safety evaluated using fluoroscopy, echocardiography and histopathology. In a healthy ovine model, Mitra-Spacer implantation was feasible and safe up to 90 days, with 7 of 8 animals surviving and no evidence of structural damage or cardiac deterioration.