Key result
Emergency manual internal cardioversion at 40 J using a left ventricular defibrillator lead successfully terminated refractory ventricular fibrillation after external cardioversion failed.
Case Report (n=1)
This case report demonstrates the successful use of a left ventricular defibrillator lead for internal cardioversion in a patient with spontaneous VF refractory to external cardioversion during CRT-D implantation.
Answers are grounded in this paper and cite their sources.
May offer a rescue option for refractory VF during device implantation; leaves open generalizability and need for prospective validation.
A 63-year-old male with ischaemic cardiomyopathy, New York Heart Association III and an ejection fraction of 25% had a bi-ventricular implantable cardioverter defibrillator (CRT-D) implant. The procedure was performed under general anaesthesia and during the placement of the defibrillator lead he went into spontaneous ventricular fibrillation (VF). External defibrillation pads were placed in the anterior-apex position and cardioversion with the Lifepak 9 (Medtronic) at 360 J was attempted twice but failed to revert to sinus rhythm. Cardiac pulmonary resuscitation (CPR) was commenced immediately with advanced cardiac life support measures but there was no return of spontaneous circulation for 35 min. He was given more than 15 external shocks at 360 J in between CPR, which were unable to terminate the VF. Finally insertion of a defibrillator lead was attempted during CPR and we managed to place the lead into the left ventricle via a subclavian artery puncture (Figure 1). The defibrillator lead connected to Unify QuadraTM (St Jude Medical). As there was undersensing by the device, emergency manual internal cardioversion at 40 J was performed but was only successful on the second attempt. Anterior-posterior view showing a defibrillator lead in the LV and RV. We report our experience and believe this is the first case of spontaneous VF that is refractory to external cardioversion during CRT-D implant. The full-length version of this report can be viewed at: http://www.escardio.org/communities/EHRA/publications/ep-case-reports/Documents/refractory-ventricular.pdf
No takes yet. Share an insight, caveat, or question.
Chow et al. (2013) conducted a case report in Refractory ventricular fibrillation during CRT-D implantation (n=1). Left ventricular defibrillator lead insertion and internal cardioversion vs. External cardioversion was evaluated on Successful cardioversion to sinus rhythm. Emergency manual internal cardioversion at 40 J using a left ventricular defibrillator lead successfully terminated refractory ventricular fibrillation after external cardioversion failed.
Synapse has enriched 10 related papers. Consider them for comparative context: