Answers are grounded in this paper and cite their sources.
The design and construction of a completely screened variable mutual inductance of high stability are described. The instrument has a range of 10,000 μH., and changes of 0·01 μH. can be read. Double screening renders the properties of the instrument independent of the circuit in which it is used. A novel method of adjustment of the sections is described, together with a novel form of inductance slide-wire. The instrument is compensated for phase defect, which even at 3000 cycles per second is only 0·0005 radian at the maximum setting. The variation of mutual inductance with frequency is also low, being less than 4 parts in 104 at 1000 cycles per second. The method of calibration is described, the limits of error being ±0·02 μH. throughout the range. The instrument is admirably suited for high precision a.c. bridge measurements of all kinds.
No takes yet. Share an insight, caveat, or question.
Astbury et al. (1938) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: