Key result
Fetal 3D cine MRI yields ~1-point lower diagnostic quality than 2D with only ~27% success.
Why the study?
This study aimed to explore a fetal 3D cardiovascular cine MRI acquisition using radial sampling and compressed-sensing reconstruction, and to compare its image quality and scan time with conventional multislice 2D imaging.
Does fetal 3D cardiovascular cine MRI using radial sampling and compressed-sensing reconstruction provide comparable cardiac diagnostic quality and quantitative cardiac measurements with shorter scan and planning times compared to conventional multislice 2D cine MRI?
Observational (n=26)
Observers were blinded to whether images were 2D or 3D acquisitions, had no information about the subject, and the images were presented in a randomized order
Does fetal 3D cardiovascular cine MRI using radial sampling and compressed-sensing reconstruction provide comparable cardiac diagnostic quality and quantitative cardiac measurements with shorter scan and planning times compared to conventional multislice 2D cine MRI?
This proof-of-concept study demonstrates that whole-heart fetal 3D radial cine MRI with compressed sensing can simplify scan planning to a single localizer and enable retrospective multiplanar cardiac views with accurate volumetric measurements, though fetal motion compensation is needed to improve its 27% acquisition success rate.
Answers are grounded in this paper and cite their sources.
Low success (27%) and moderate quality preclude clinical use of fetal 3D radial cine MRI; leaves open feasibility after motion compensation advances.
PURPOSE: To explore a fetal 3D cardiovascular cine acquisition using a radial image acquisition and compressed-sensing reconstruction and compare image quality and scan time with conventional multislice 2D imaging. METHODS: Volumetric fetal cardiac data were acquired in 26 volunteers using a radial 3D balanced SSFP pulse sequence. Cardiac gating was performed using a Doppler ultrasound device. Images were reconstructed using a parallel-imaging and compressed-sensing algorithm. Multiplanar reformatting to standard cardiac views was performed before image analysis. Clinical 2D images were used for comparison. Qualitative and quantitative image evaluation were performed by two experienced observers (scale: 1-4). Volumes, mass, and function were assessed. RESULTS: Average scan time for the 3D imaging was 6 min, including one localizer. A 2D imaging stack covering the entire heart including localizer sequences took at least 6.5 min, depending on planning complexity. The 3D acquisition was successful in 7 of 26 subjects (27%). Overall image contrast and perceived resolution were lower in the 3D images. Nonetheless, the 3D images had, on average, a moderate cardiac diagnostic quality (median [range]: 3 [1-4]). Standard clinical 2D acquisitions had a high cardiac diagnostic quality (median [range]: 4 [3, 4]). Cardiac measurements were not different between 2D and 3D images (all p > 0.16). CONCLUSION: The presented free-breathing whole-heart fetal 3D radial cine MRI acquisition and reconstruction method enables retrospective visualization of all cardiac views while keeping examination times short. This proof-of-concept work produced images with diagnostic quality, while at the same time reducing the planning complexity to a single localizer.
No takes yet. Share an insight, caveat, or question.
Piek et al. (2022) conducted an observational in Fetal cardiac assessment in pregnant women (including congenital heart defects) (n=26). Fetal 3D cardiovascular cine MRI acquisition using radial sampling and compressed-sensing reconstruction vs. Conventional multislice 2D Cartesian balanced SSFP cine MRI was evaluated on Cardiac diagnostic quality score (scale 1-4) (p=0.009). Fetal 3D radial cine MRI succeeded in 27% (7/26) of subjects with moderate diagnostic quality and comparable cardiac measurements.
Synapse has enriched 10 related papers. Consider them for comparative context: