This is the authors' abstract. We don't add key points for this paper.
We present a four beam ratiometric setup for an integrating sphere based gas cell, which can correct for changes in pathlength due to sphere wall contamination. This allows for the gas absorption coefficient to be determined continuously without needing to recalibrate the setup. We demonstrate the technique experimentally, measuring methane gas at 1651nm. For example, contamination covering 1.2% of the sphere wall resulted in an uncompensated error in gas absorption coefficient of ≈41%. With the ratiometric scheme, this error was reduced to ≈2%. Potential limitations of the technique, due to subsequent deviations from mathematical assumptions are discussed, including severe sphere window contamination.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Bergin et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: