Why the study?
Does noninvasive measurement using photoelectric plethysmography accurately determine the volume elastic modulus in finger arteries compared to pulse wave propagation velocity?
Does noninvasive measurement using photoelectric plethysmography accurately determine the volume elastic modulus in finger arteries compared to pulse wave propagation velocity?
This noninvasive photoelectric plethysmography method provides a valid measurement of arterial volume elastic modulus, correlating well with pulse wave propagation velocity.
May support noninvasive arterial elasticity assessment; leaves open clinical adoption pending larger validation studies.
A method for the nonivasive measurement of the volume elastic modulus (Ev) in human finger arteries was developed. The volume change ratio and pulse pressure in the arteries were simultaneously determined by a transmittance type infrared photoelectric plethysmograph and a volume oscillometric sphygmomanometer using an occlusive cuff. The Ev values could be obtained at a desired transmural pressure level of the arteries which was controlled by the cuff. A clear difference was revealed between the Ev values obtained in the fingers of two male subjects of 33 and 65 years old. The values determined by this method were compared to those calculated from pulse wave propagation velocity (Evp) obtained simultaneously in the fingers of 11 subjects. The linear regression equation was Ev = 1.31 Evp-0.117 with a correlation coefficient r = 0.893 (n = 39). These results suggest that this should be a useful and noninvasive method of evaluating changes in arterial elasticity accompanying aging and perhaps influenced by accompanying hypertension or arteriosclerosis.
No takes yet. Share an insight, caveat, or question.
Shimazu et al. (1986) studied this question.