Key result
Implementation of the rapid Xpert Norovirus assay as a 24/7 service demonstrated 100% sensitivity and 93% specificity while significantly reducing median turn-around time from 22 hours to 2.4 hours.
Why the study?
Does the Xpert Norovirus assay improve diagnostic turn-around time and provide accurate detection of norovirus in patients with acute gastroenteritis?
Observational (n=276)
Yes
Does the Xpert Norovirus assay improve diagnostic turn-around time and provide accurate detection of norovirus in patients with acute gastroenteritis?
Absolute Event Rate: 2.4% vs 22%
p-value: p=<0.001
Implementation of the rapid Xpert Norovirus assay as a 24/7 service provided excellent diagnostic accuracy and significantly reduced turn-around time for norovirus detection.
May expedite infection control via faster diagnosis; supports assay use but leaves outcome benefits open.
Noroviruses are a leading cause of epidemic and sporadic cases of acute gastroenteritis worldwide. The rapid diagnosis of norovirus infection is important for prompt infection control measures and may reduce the need for additional diagnostic testing. Here we evaluated the performance of the rapid Xpert Norovirus assay, and assessed the turn-around time (TAT) before and after the implementation of the analysis as a 24/7 service at all the three hospitals in Jönköping County, Sweden. We describe the implementation process which was performed in two steps during 2014. A total number of 276 clinical samples (stool and vomitus) from patients with symptoms of acute gastroenteritis were included in 2014-2015. The samples were analysed with the Xpert Norovirus assay and the already existing routine method: an in-house reverse transcription real-time PCR. Samples showing discrepant results with the two assays were further analysed by a third PCR method. The Xpert Norovirus assay performed well with a sensitivity of 100% and a specificity of 93% compared to the gold standard (defined as the result obtained by at least two of the three PCR methods). The median TAT decreased from 22 hours in 2013 to 2.4 hours in 2015 (p<0.001). We conclude that the performance of the Xpert Norovirus assay was excellent, and that the implementation of the analysis as a 24/7 service at all three hospitals in the county has greatly reduced the time to diagnosis which is beneficial for both patients and healthcare providers.
No takes yet. Share an insight, caveat, or question.
Henningsson et al. (2017) conducted an observational in Acute norovirus-associated gastroenteritis (n=276). Xpert Norovirus assay implemented as a 24/7 service vs. Routine in-house PCR method (historical control for turn-around time) was evaluated on Median turn-around time (TAT) in hours (p=<0.001). Implementation of the rapid Xpert Norovirus assay as a 24/7 service demonstrated 100% sensitivity and 93% specificity while significantly reducing median turn-around time from 22 hours to 2.4 hours.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: