Key result
Implantable cardioverter defibrillators in patients with Brugada syndrome resulted in multiple inappropriate shocks in 66.7% (2 of 3) of patients due to dynamic T wave oversensing.
Case Report (n=7)
In Brugada syndrome, dynamic variations in right ventricular endocardial electrograms can cause T wave oversensing and inappropriate ICD shocks, which may be prevented by implanting a left ventricular epicardial lead.
May prompt caution with endocardial ICD leads in Brugada syndrome due to oversensing risk; leaves open alternative configurations for prospective evaluation.
INTRODUCTION: Dynamic variations in electrophysiologic phenomena inherent to the Brugada syndrome may complicate therapy with implantable cardioverter defibrillators (ICDs). METHODS AND RESULTS: Between 1997 and 1999, 3 of 7 patients with Brugada syndrome (1 man and 2 women, mean age 42 years) received an ICD. During follow-up, 2 patients experienced multiple inappropriate shocks. Simultaneously with dynamic changes in the surface ECG, endocardial ECGs revealed a dynamic decrease in the right ventricular R wave and an increase in the corresponding T wave, resulting in T wave oversensing. With ajmaline administration, these dynamic changes in endocardial signals were reproducible at different right ventricular sites, whereas left ventricular epicardial signals remained stable. Incremental AAI pacing and exercise stress testing resulted in similar changes in right ventricular endocardial signals, but normalization of the surface ECG apart from progressively increasing S waves in leads II, V5, and V6. Orciprenaline administration had no effect on ECG phenomena. After implantation of a left ventricular epicardial lead for sensing and pacing, no inappropriate tachycardia detection recurred. CONCLUSION: These findings demonstrate that, in Brugada syndrome, spontaneous or ajmaline-induced changes in the surface ECG may be paralleled by significant variations in the right ventricular endocardial electrogram that may result in ICD malfunction. Implantation of a left ventricular epicardial lead for sensing and pacing may be the ultimate successful approach in certain patients. To assure proper ICD function, ajmaline testing during ICD implantation appears to be helpful.
No takes yet. Share an insight, caveat, or question.
Stix et al. (2000) conducted a case report in Brugada syndrome (n=7). Implantable cardioverter defibrillator (ICD) was evaluated on Inappropriate shocks. Implantable cardioverter defibrillators in patients with Brugada syndrome resulted in multiple inappropriate shocks in 66.7% (2 of 3) of patients due to dynamic T wave oversensing.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: