Answers are grounded in this paper and cite their sources.
The classic Severinghaus p CO 2 microelectrodes have been greatly improved. The drift is reduced from ∼10 to 0.4 mV/h, and the lifetime has been extended from hours or days to one month. The detection limit has been lowered to 200 μatm, and the t 98 for low p CO 2 response is 2 min. Therefore, the microelectrodes can be extended to low p CO 2 measurements. In addition, the microelectrodes can be fabricated more easily. The major problems with the classic p CO 2 microelectrodes are caused by Pyrex and the reference electrode used for constructing the microelectrodes. Pyrex can release chemicals into the internal bicarbonate solution and cause serious drift and a short lifetime. Pyrex has been replaced by aluminosilicate glass, and appropriate Ag/AgCl reference electrodes have been used. The enzyme carbonic anhydrase has been added to the internal bicarbonate solution to accelerate the hydration reaction, which greatly enhances the response rate and improves the detection limit. The internal pH glass microelectrodes have been replaced by pH liquid membrane microelectrodes, which makes fabrication easier and further improves response rate. The theory of the electrode's response characteristics at low p CO 2 is presented as well as a discussion on the effects of glass materials, the enzyme and pH electrodes.
No takes yet. Share an insight, caveat, or question.
Zhao et al. (1997) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: