Key result
Volumetric measurement of carotid-femoral pulse wave velocity using Vicorder was only moderately correlated with tonometric measurement using SphygmoCor (R = 0.50), indicating the values are not interchangeable.
Why the study?
Does volumetric measurement of pulse wave velocity (Vicorder) provide interchangeable results with tonometric measurement (SphygmoCor) in children?
Cross-Sectional (n=156)
No
Does volumetric measurement of pulse wave velocity (Vicorder) provide interchangeable results with tonometric measurement (SphygmoCor) in children?
Effect estimate: R = 0.50
p-value: p=<0.001
Volumetric measurement of pulse wave velocity is reproducible and easy to use in children, but values are not interchangeable with tonometric methods.
Volumetric PWV measurement may be feasible in children; leaves open whether values are interchangeable with tonometry for risk stratification.
BACKGROUND: Pulse wave velocity (PWV), a measure of arterial stiffness strongly predictive of cardiovascular risk in adults, is usually measured by sequential ECG-referenced carotid and femoral tonometry. A simplified technique, more suitable for use in children, employs simultaneous volumetric recording from a sensor applied over the carotid artery and a cuff applied over the femoral artery or arm and thigh pressure cuffs applied over the brachial and femoral arteries. The purpose of this study was to compare PWV computed over the carotid-femoral path (PWVcf) with that over the brachial-femoral path (PWVbf) using a volumetric system (Vicorder) and to compare values of PWVcf obtained by the volumetric and a tonometric method (SphygmoCor) in children. METHOD: Vicorder PWVcf and PWVbf were compared in 156 children (3-18 years, 110 with chronic kidney disease), and PWVcf by Vicorder was compared to PWVcf by SphygmoCor in a subset of 122 patients. RESULTS: PWVcf by Vicorder was moderately correlated with PWVcf by SphygmoCor (R = 0.50, P < 0.000). PWVbf and PWVcf Vicorder were more closely correlated (R = 0.75, P < 0.0001), but with a significant systematic difference. Applying a correction factor to PWVbf measurements gave results similar to those obtained over the carotid-femoral path. Within-patient coefficients of variation for repeated measures were 5.9, 7.8, and 8.5% for PWVbf (Vicorder), PWVcf (Vicorder) and PWVcf (SphygmoCor), respectively. All PWV values showed a similar relation to age. CONCLUSION: Volumetric methods appear reproducible and are easy to use in children, but values obtained by Vicorder and SphygmoCor are not interchangeable even when measured over the same pathway.
No takes yet. Share an insight, caveat, or question.
Keehn et al. (2014) conducted a cross-sectional in Chronic kidney disease and hypertension (n=156). Volumetric pulse wave velocity measurement (Vicorder) vs. Tonometric pulse wave velocity measurement (SphygmoCor) was evaluated on Correlation between carotid-femoral pulse wave velocity measured by Vicorder and SphygmoCor (R = 0.50, p=<0.001). Volumetric measurement of carotid-femoral pulse wave velocity using Vicorder was only moderately correlated with tonometric measurement using SphygmoCor (R = 0.50), indicating the values are not interchangeable.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: