Key result
A semiautomatic portable blood pressure recorder showed good correlation with the ordinary Riva-Rocci method for systolic and diastolic measurements.
Why the study?
Does a portable semiautomatic blood pressure recorder provide measurements correlated with the ordinary Riva-Rocci method?
Does a portable semiautomatic blood pressure recorder provide measurements correlated with the ordinary Riva-Rocci method?
A newly described portable semiautomatic blood pressure recorder shows good correlation with the standard Riva-Rocci method for both systolic and diastolic measurements.
Answers are grounded in this paper and cite their sources.
May support portable device validation; leaves open outcome trials before clinical adoption.
A semiautomatic portable recorder for measurement of blood pressure, heart rate, and time of observation is described. The instrument is based on the Riva-Rocci method with a microphone in the cuff. The Korotkoff sounds and the cuff pressure are converted to electric impulses and recorded on a tape recorder. The read-out system is simple. The correlation between simultaneous measurements of blood pressure by the instrument and by the ordinary Riva-Rocci method is good (r systolic=0.92 and r diastolic=0.84).
No takes yet. Share an insight, caveat, or question.
Lund‐Johansen et al. (1972) studied this question. Semiautomatic portable blood pressure recorder vs. Ordinary Riva-Rocci method was evaluated on Correlation between simultaneous measurements of blood pressure (r systolic=0.92 and r diastolic=0.84). A semiautomatic portable blood pressure recorder showed good correlation with the ordinary Riva-Rocci method for systolic and diastolic measurements.