Why the study?
Does a new automatic ambulatory blood pressure recorder provide reliable and reproducible measurements compared to a sphygmomanometer in normotensive and hypertensive subjects?
Does a new automatic ambulatory blood pressure recorder provide reliable and reproducible measurements compared to a sphygmomanometer in normotensive and hypertensive subjects?
The novel ambulatory automatic blood pressure monitoring system provides reliable and reproducible measurements at rest over 24 hours, though it is not suitable for use during exercise stress testing.
Answers are grounded in this paper and cite their sources.
May support resting ambulatory BP monitoring with the device; leaves open accuracy during exercise.
Reliability and reproducibility of a new automatic ambulatory blood pressure recorder were tested on 31 subjects (18 normotensives and 13 hypertensives). Of 3688 recordings, 1424 measurements were cross-checked with a sphygmomanometer in standing or sitting position at the beginning and at the end of a twenty-four hour monitoring period. In the normotensive subjects cross-checkings were performed also during a treadmill exercise stress test. Analysis of variance did not indicate statistical differences between the two types of measurement (p greater than 0.09 for systolic and p greater than 0.62 for diastolic), and a significant linear correlation was found at rest (r = 0.99, p less than 0.001 for systolic and r = 0.98, p less than 0.001 for diastolic pressure). During the effort test the device was unable to correctly measure blood pressure values after the first step. During the twenty-four hour monitoring period only 1.1% of the preprogrammed measurements were lost for each subject. The recorder seems to be a suitable and reliable tool for automatic blood pressure monitoring.
No takes yet. Share an insight, caveat, or question.
Modesti et al. (1990) studied this question.