Evaluation of the Reproducibility and Accuracy of Ambulatory Blood Pressure Monitoring Using the Takeda TM-2420 Automated Blood Pressure Monitor

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Does the Takeda TM-2420 automated blood pressure monitor provide reproducible and accurate ambulatory blood pressure measurements in normotensive subjects?

Population

8 normotensive subjects

Comparison

Ambulatory blood pressure monitoring using… vs Standard sphygmomanometer measurement

Design

Other

Follow-up

3 separate days

Authors

JKJohn J. KellyMayo ClinicSHStephen N. HunyorNorthern Sydney Local Health DistrictKHK. Y. HoGarvan Institute of Medical Research

Discussion

Loading...

Member takes

Implication

Supports mean ambulatory BP reproducibility with Takeda TM-2420; leaves open variability assessment before wider use.

Structured PICO

Does the Takeda TM-2420 automated blood pressure monitor provide reproducible and accurate ambulatory blood pressure measurements in normotensive subjects?

P
Population
8 normotensive subjects
I
Intervention
Ambulatory blood pressure monitoring using Takeda Medical TM-2420 automated blood pressure monitor over three separate days of recording
C
Comparator
Standard sphygmomanometer measurement
O
Outcome
Reproducibility (coefficients of variation) of mean 24-hour systolic and diastolic blood pressure and accuracy (bias) compared to standard sphygmomanometersurrogate

The Takeda TM-2420 automated blood pressure monitor provides reproducible estimates of mean systolic and diastolic blood pressure, though blood pressure variability is poorly reproducible.

Cite This Study

Kelly et al. (1993) studied this question.

synapsesocial.com/papers/6a81503ddd9b32ddc29cd291https://doi.org/10.3109/10641969309032934
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Bench and ambulatory field evaluation of the A & D TM-2420 automated sphygmomanometer1990 · 19 citations
  2. 2Reproducibility of non-invasive and intra-arterial blood pressure monitoring: implications for studies on antihypertensive treatment1991 · 159 citations
  3. 3Ambulant Blood Pressure: Reproducibility and the Assessment of Interventions1980 · 45 citations
  4. 4Reproducibility of the Whole-Day Blood Pressure Pattern in Essential Hypertension1982 · 38 citations
  5. 5Continuous vs intermittent blood pressure measurements in estimating 24-hour average blood pressure.1983 · 239 citations