Key result
Transit-time ultrasound showed excellent agreement with Doppler ultrasound for measuring flow in aortocoronary bypass grafts in vivo.
Why the study?
Does the ultrasound transit-time method accurately measure flow in aortocoronary bypass grafts compared to the Doppler ultrasound method?
Observational (n=9)
Does the ultrasound transit-time method accurately measure flow in aortocoronary bypass grafts compared to the Doppler ultrasound method?
The ultrasound transit-time method demonstrates excellent accuracy and agreement with the Doppler method for measuring flow in aortocoronary bypass grafts both in vitro and in vivo.
Answers are grounded in this paper and cite their sources.
May support transit-time ultrasound for graft flow assessment; leaves open larger prospective validation before routine adoption.
To evaluate the accuracy of flow measurements in aortocoronary bypass grafts with the ultrasound transit-time method, an in vitro and in vivo comparison was carried out. The in vitro comparison with evaluation against both true flow and the ultrasound Doppler method, was carried out with a fresh saphenous vein mounted in a pulsatile flow rig. The two flow probes were placed on the graft 4-5 cm apart to avoid acoustic interference, and blood was pumped through the system at different flow rates. The comparison between the methods showed excellent agreement with a linear correlation coefficient of 0.996, and a mean error of -2.9 ml/min with limits of agreement +/- 13.1 ml/min (+/- 2 SD = 95% of measured differences between these limits). However, against true flow, both methods overestimated flow slightly with mean error 4.4 and 7.3 ml/min for the transit-time and Doppler, respectively. Both methods showed excellent correlation with true flow (correlation coefficient 0.998 for the transit-time and 0.997 for the Doppler method). The in vivo accuracy was evaluated by comparison of the two methods in 9 patients. The two probes were placed on the same saphenous vein grafts 4-5 cm apart, and a total of 34 measurements in 17 grafts were carried out including measurements at baseline and after papaverine injection. The correlation coefficient was 0.990 and linear regression analysis gave the equation: Transit-time flow = 1.00 x Doppler flow + 1.3. In terms of flow, the mean error was 1.5 ml with limits of agreement +/- 17.2 ml.(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
Matre et al. (1994) conducted an observational in Aortocoronary bypass grafts during cardiac surgery (n=9). Transit-time ultrasound method vs. Doppler ultrasound method was evaluated on Agreement of flow measurements in vivo (Correlation coefficient 0.990). Transit-time ultrasound showed excellent agreement with Doppler ultrasound for measuring flow in aortocoronary bypass grafts in vivo.