Answers are grounded in this paper and cite their sources.
A capillary zone electrophoresis (CZE) method was developed for the rapid analysis of charge heterogeneity of immunoglobulin G (IgG) monoclonal antibodies (mAbs). The separation was carried out in a short, dynamically coated fused-silica capillary. A number of separation parameters were investigated and optimized, including pH, concentration of the separation buffer (ε-amino caproic acid), concentration of the triethylenetetramine (TETA) dynamic coating, the capillary internal diameter and the field strength used for the separation. The effects of between-run flushing of the capillary and the data acquisition rate were also evaluated. Under the optimized conditions, a fast (<5 min), selective and reproducible separation of mAb charge variants was achieved under a very high electric field strength (1000 V/cm). This method also requires only a short conditioning of the capillary, with between-run conditioning completed within 2 min. The method was evaluated for specificity, sensitivity, linearity, accuracy and precision. The same separation conditions were applied to the rapid separation (2-5 min) of charge variants of multiple monoclonal antibodies with pI in the range of 7.0-9.5. Compared with other existing methods for charge variants analysis, this method has several advantages including a short run time, rapid capillary conditioning and simple sample preparation.
No takes yet. Share an insight, caveat, or question.
He et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: